EasyManua.ls Logo

Keysight Technologies 34420A

Keysight Technologies 34420A
189 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Front-Panel Operation
50 Keysight 34420A Service Guide
G: CALibration MENU
[1]
[1] The commands enclosed in square brackets ( [ ] ) are “hidden” unless the meter is UNSECURED for
calibration.
1: SECURED ) [1: UNSECURED] ) [2:CALIBRATE] ) [3: CHART ZERO] ) [4 : CHART GAIN] )
[5: INJECTED I] ) 6:CAL COUNT ) 7:MESSAGE
1: SECURED The meter is secured against calibration; enter code to
unsecure.
1: UNSECURED The meter is unsecured for calibration; enter code to
secure.
2: CALIBRATE Performs a calibration of meter; must be UNSECURED.
3: CHART ZERO Performs a calibration of the chart recorder output zero
level.
4: CHART GAIN Performs a calibration of the chart recorder output gain.
5: INJECTED I Performs a calibration to minimize the injected current.
6: CAL COUNT Reads the total number of times the meter has been
calibrated.
7: MESSAGE Reads the first 11 characters of a calibration string, if any,
entered from the remote interface.

Table of Contents

Related product manuals