Calibration Procedures 4
Keysight 34970A/34972A Service Guide 111
17 Ch 31 & 41 P3 P3 P4 P4 ________Ohms 2.00 W K301
18 Ch 32 & 42 P3 P3 P4 P4 ________Ohms 2.00 W K302
19 Ch 33 & 43 P3 P3 P4 P4 ________Ohms 2.00 W K303
20 Ch 34 & 44 P3 P3 P4 P4 ________Ohms 2.00 W K304
21 Ch 35 & 45 P3 P3 P4 P4 ________Ohms 2.00 W K305
22 Ch 36 & 46 P3 P3 P4 P4 ________Ohms 2.00 W K306
23 Ch 37 & 47 P3 P3 P4 P4 ________Ohms 2.00 W K307
24 Ch 38 & 48 P3 P3 P4 P4 ________Ohms 2.00 W K308
25 Ch 41 & 11 P4 P4 P1 P1 ________Ohms 2.00 W K401
26 Ch 42 & 12 P4 P4 P1 P1 ________Ohms 2.00 W K402
27 Ch 43 & 13 P4 P4 P1 P1 ________Ohms 2.00 W K403
28 Ch 44 & 14 P4 P4 P1 P1 ________Ohms 2.00 W K404
29 Ch 45 & 15 P4 P4 P1 P1 ________Ohms 2.00 W K405
30 Ch 46 & 16 P4 P4 P1 P1 ________Ohms 2.00 W K406
31 Ch 47 & 17 P4 P4 P1 P1 ________Ohms 2.00 W K407
32 Ch 48 & 18 P4 P4 P1 P1 ________Ohms 2.00 W K408
[a] Only the channel currently under test should be closed at one time. All other channels should be open.
Test #
Channel
Closed
[a]
External DMM Ohmmeter Connections
Measured Value
Test
Limit
Relay Measured
HI LO HI Sense LO Sense