EasyManua.ls Logo

Keysight Technologies 34972A - Operating Checklist

Keysight Technologies 34972A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service 6
Keysight 34970A/34972A Service Guide 171
Operating Checklist
Before returning your instrument to Keysight Technologies for service or repair,
check the following items:
Is the instrument inoperative?
Verify that the AC power cord is connected to the instrument.
Verify that the front-panel On/Standby switch has been pushed.
Verify that the power-line fuse is installed and not open (see page 41). The
instrument is shipped from the factory with a 500 mAT, 250 V fuse. This is the
correct fuse for all line voltages.
Verify the power-line voltage setting. See “If the Instrument Does Not Turn On”
on page 41.
Does the instrument fail self-test?
Verify that the correct power-line voltage is selected. See “If the Instrument
Does Not Turn On” on page 41.
Remove all input connections to the instrument. Errors may be induced by AC
signals present on the input wiring during a self-test. Long test leads can act
as an antenna causing pick-up of AC signals.
Is the Current measurement function inoperative?
Verify the input protection fuses on the 34901A Multiplexer Module. If
necessary, replace with a 1.5 A, 250 V NTD fuse.
NOTE
The Current input is only available on channels 21 and 22 of the 34901A module.

Table of Contents

Other manuals for Keysight Technologies 34972A

Related product manuals