4 Calibration Procedures
122 Keysight 34970A/34972A Service Guide
34907A Analog Output
Analog Output Verification Test
This procedure is used to check the calibration of the analog outputs on the
34907A Multifunction Module. Install the module in slot 200. Verification checks
are performed only for those output values with unique calibration constants.
1 Make connections to analog output channels as shown below.