Keysight E5260/E5270 Programming Guide, Edition 4 4- 31
Command Reference
Command Reference
Remarks In the limit search mode, if search cannot find the search target and the following
two conditions are satisfied, the E5260/E5270 repeats the binary search between the
last source value and the source start value.
• target is between the data at source start value and the last measurement data.
• target is between the data at source stop value and the data at:
source value = | stop − start | / 2.
If the search cannot find the search target and the following two conditions are
satisfied, the E5260/E5270 repeats the binary search between the last source value
and the source stop value.
• target is between the data at source stop value and the last measurement data.
• target is between the data at source start value and the data at:
source value = | stop − start | / 2.