1- 12 Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Getting Started
To Pause Command Execution
To pause command execution until the specified wait time has elapsed, send the PA
command.
Example OUTPUT @E5270;"PA 5"
If this command is sent, the E5260/E5270 waits 5 seconds before executing the next
command.
To Start Measurement
To start measurement other than the high speed spot measurement, send the XE
command.
Example OUTPUT @E5270;"XE"
This starts the measurement specified by the MM command.
For the high speed spot measurement, refer to “To Perform High Speed Spot
Measurement” on page 1-18.
To Force 0 V
To force 0 V immediately, send the DZ command. The E5260/E5270 memorizes the
present source output settings of the specified channel, and changes the specified
channel output to 0 V. If you do not specify the channel, the DZ command function
is effective for all channels.
Example OUTPUT @E5270;"DZ 1"
If this command is sent, the E5260/E5270 memorizes the current settings of channel
1 (the module installed in slot 1 of the E5260/E5270), and changes channel
1 output
to 0 V.
To restore the settings stored by the DZ command, send the RZ command. For
details, refer to
Chapter 4, “Command Reference.”

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals