Keysight E5260/E5270 Programming Guide, Edition 4 4- 39
Command Reference
Command Reference
BSV
The BSV command specifies and sets the voltage search source for the binary
search measurement. This command setting is cleared by the BSI command. After
search stops, the search channel forces the value specified by the BSM command.
Syntax BSV chnum,range,start,stop[,Icomp]
Parameters chnum : Search source channel number. The value must be slot number where
the module has been installed. Integer expression. 1 to 8 (1 and 5 are
not available for HPSMU). See
Table 4-1 on page 4-9.
range : Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See
Table 4-4 on page 4-13.
start, stop : Search start or stop voltage (in V). Numeric expression. See Table 4-6
on page 4-14. The start and stop parameters must have different values.
0 to ±100, or 0 to ±200 for E5280B/E5290A
Icomp : Current compliance value (in A). Numeric expression. See Table 4-6 on
page 4-14. If you do not specify Icomp, the previous value is set. Zero
amps (0 A) is not allowed for Icomp.
Example
Statements
OUTPUT @E5270;"BSV 1,0,0,20,1E-6"