4- 14 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Parameters
Table 4-6 Voltage Source Setup Parameters for
DV/TDV/BDV/WV/WSV/WNX/PV/PWV/LSV/BSV Commands
Output
range
(actually
used)
Setting
resolution
in V
voltage, start,
stop, base, or
pulse in
V
Maximum Icomp value in A
E5280B E5281B E5287A E5290A E5291A
0.5 V 25E-6 0 to ± 0.5 NA ±100E-3 ±100E-3 NA NA
2 V 100E-6 0 to ± 2 ±1 ±100E-3 ±100E-3 ±1 ±200E-3
5 V 250E-6 0 to ± 5 NA ±100E-3 ±100E-3 NA NA
20 V 1E-3 0 to ± 20 ±1 ±100E-3 ±100E-3 ±1 ±200E-3
40 V 2E-3 0 to ± 20 ±500E-3 ±100E-3 ±100E-3 ±500E-3 ±200E-3
to ± 40 ±50E-3 ±50E-3 ±50E-3
100 V 5E-3 0 to ± 20 ±125E-3 ±100E-3 ±100E-3 ±125E-3 ±200E-3
to ± 40 ±50E-3 ±50E-3 ±50E-3
to ± 100 ±20E-3 ±20E-3 ±20E-3
200 V 10E-3 0 to ± 200 ±50E-3 NA NA ±50E-3 NA