1- 46 Keysight E5260/E5270 Programming Guide, Edition 4
Programming Basics
Programming Tips
To Perform Quasi-Sampling Measurement
The following setup enables to perform a quasi-sampling measurement. Then the
sampling interval will be sum of delay time and step delay time.
Sets the sweep measurement mode (MM 2 or MM 16).
Sweep start value = Sweep stop value (for WI, WV, or WNX).
Sets hold time, delay time, and step delay time (WT).
To Interrupt Command Execution
The E5260/E5270 executes commands in the received order. However, only the
following commands can interrupt the command execution.
Table 1-5 Interrupt Commands
Voltage or current
Measurement trigger (e.g. XE)
Hold
Start value, Stop value
Previous value
: Measurement
Delay Step delay
Sampling interval = Delay + Step delay
Command Description
AV Changes the number of averaging samples during the
measurement.
AIT Changes the integration time during the measurement.
AB Aborts the command execution.
*RST Resets the E5260/E5270 during the command execution.
XE If the E5260/E5270 has been set to the wait status by the PA or
PAX command, the XE command can be used to release the
wait status. For details, refer to
Chapter 4, “Command
Reference.”

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals