Keysight E5260/E5270 Programming Guide, Edition 4 4- 45
Command Reference
Command Reference
CMM
The CMM command sets the SMU measurement operation mode. This command is
not available for the high speed spot measurement.
Syntax CMM chnum,mode
Parameters chnum : Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
Table 4-1 on page 4-9.
mode : SMU measurement operation mode. Integer expression.
0: Compliance side measurement (initial setting).
1: SMU always performs current measurement.
2: SMU always performs voltage measurement.
3: Force side measurement.
If mode=0, SMU measures current when it forces voltage, or measures voltage when
it forces current.
If mode=3, SMU measures current when it forces current, or measures voltage when
it forces voltage.
The mode setting is kept until the mode is changed by this command. If you want to
return it to the initial setting, enter the CMM command with mode=0.
Example
Statements
OUTPUT @E5270;"CMM 1,1"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals