BSVM
The BSVM command selects the data output mode for the binary search
measurement.
Syntax BSVM mode
Parameters mode : Data output mode. Integer expression.
0 : Returns Data_search only (initial setting).
1 : Returns Data_search and Data_sense.
where
Data_search is the value forced by the search output channel set by the
BSV or BSI command.
Data_sense is the value measured by the search monitor channel set by
the BGI or BGV command.
For data output format, refer to “Data Output Format” on page 1-22.
Example
Statements
OUTPUT @E5270;"BSVM 1"