EasyManuals Logo

Keysight Technologies N5183B MXG User Manual

Keysight Technologies N5183B MXG
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #366 background imageLoading...
Page #366 background image
348 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Using Avionics VOR/ILS Softkeys for N5171B/72B and N5181B/82B with Option 302
Using VOR [VHF Omnidirectional Range] Softkeys
To set the VOR bearing direction as From or To
1. Press Aux Fctn > Avionics > VOR
2. Press Bearing
3. Press Direction
(Default selection is From.)
4. Select either From | To
From - In the From convention, the VOR transmitter beacon is made the
reference point and the Bearing Angle is between local magnetic-North and
the beacon-to-aircraft radial line (RL).
To - In the To convention, the aircraft is made the reference point and the
Bearing Angle is between local magnetic-North and the aircraft-to-beacon
radial line (RL).
RL
(To)
= 180° – RL
(From)
To set the frequency of the reference signal (REF Freq) and the
variable phase signal (VAR Freq)
1. Press Aux Fctn > Avionics > VOR
2. Press REF/VAR
3. Press REF/VAR Freq
(Default <value> is 30 Hz.)
4. Select a <value> in Hz | kHz | MHz | GHz from 10 to 60 Hz.
Enter a value in Hz to set the frequency of the reference signal (REF Freq) and
the variable phase signal (VAR Freq).

Table of Contents

Other manuals for Keysight Technologies N5183B MXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5183B MXG and is the answer not in the manual?

Keysight Technologies N5183B MXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5183B MXG
CategoryPortable Generator
LanguageEnglish

Related product manuals