6 53210A Math, Graphing, and Data Logging
154 Keysight 53210A User’s Guide
The offset value (B) is set with the command:
CALCulate[1]:SCALe:OFFSet {<offset>|MINimum|MAXimum|DEFault}
CALCulate[1]:SCALe:OFFSet? [{MINimum|MAXimum|DEFault}] (query form)
offset values can range from
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to +1.0E+15
The default offset is 0.0 and is the value set following a reset (*RST) or
instrument preset (
SYSTem:PRESet).
Inverting the Reading (1/x) Each reading as it is taken can be inverted prior to
its use in the Mx-B equation which thus becomes M/x-B. Inversion is enabled with
the command:
CALCulate[1]:SCALe:INVert {OFF|ON}
CALCulate[1]:SCALe:INVert? (query form)
ON - inverts the reading. OFF disables reading inversion.
Following a reset (
*RST) or instrument preset (SYSTem:PRESet) reading inversion is
disabled.
When Invert x is enabled (On) the front panel reading units (Hz, sec) are turned
off. See “Assigning Reading Units” if displayed units are required.