53210A Error Messages A
Keysight 53210A User’s Guide 235
-350 Error queue overflow
An error has occurred; however, the instrument error queue is full
and the error was not recorded.
Query Errors
-410 Query IINTERRUPTED
The counter received a command before it was finished
responding to a previous query.
-420 Query UNTERMINATED
The computer attempts to read a query response from the counter
after having issued an incomplete query command.
-430 Query DEADLOCKED
The counter’s input and output buffers are full and the instrument
cannot continue.
-440
Query UNTERMINATED after indefinite
response
A query command was received in the same string after a query
(e.g. *IDN?) requesting an indefinite length response.
Instrument Errors
+100 Network Error A LAN network error has occurred or there is a possible hardware
failure. Run the instrument self-test to determine if a hardware
failure exists. If so, contact Keysight.
+110 LXI mDNS Error
+263
Not able to execute while instrument is
measuring
A command was received while a measurement is in progress
+291 Not able to recall state: it is empty
The state storage location (0-4) specified by the *RCL command is
empty
+292 State file size error Attempted to load a .sta (state) file that is too large.
+293 State file corrupt
The .sta (state) file specified does not contain valid instrument
state information.
+294 Preference file size error Attempted to load a .prf (preference) file that is too large.
+295 Preference file corrupt
The .prf (preference) file specified does not contain valid
user-preference setting information.
+301
Input termination protection relay
opened
An input signal greater than or equal to + 10Vp is present on a
channel. The input impedance is set to 1 MΩ. Removing or
reducing the signal below the damage level and pressing the
flashing (channel) key or sending INPut:PROTection:CLEar resets
the protection relay.
Table A-1 53210A Error Message Descriptions.