EasyManua.ls Logo

Keysight E36300 Series - Utilities Menu - Test; Setup; Calibration; Self Test

Keysight E36300 Series
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Utilities Menu - Test / Setup
Test / Setup provides access to calibration, self-test, as well as configures user preferences, sets the date and time,
and accesses the help:
Calibration
Calibration accesses the instrument calibration procedure.
Perform Calibration opens the calibration menu. Make sure you have run the self-test before performing any
calibration. Refer to "Calibration Adjustment Procedures" in the Service Guide for details.
The default security passcode is 0. Press Change Passcode to change the passcode using the numeric keypad. The
passcode can be set up to 9 digits.
Auto save On will automatically save the calibration data upon logout.
When Auto save is Off, press Cal Save to save the calibration data.
Reset to Factory returns the settings to factory default settings. Pressing this key sanitizes all user-accessible
instrument memory except calibration constants and reboots the instrument.
Self Test
Self Test verifies proper instrument operation. Refer to "Self-Test Procedures" in the Service Guide for details.
100
Keysight E36300 Series User's Guide

Table of Contents

Other manuals for Keysight E36300 Series

Related product manuals