EasyManuals Logo

Keysight E36300 Series User Manual

Keysight E36300 Series
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
– Setting Change: delay timer start at the end of a settings change in voltage, current, or output state. Press OCP
Start Set.
Clears and OVP and OCP Event
To clear the protection function, first remove that condition that caused the protection fault.
Press Protection Clear to clears the protection function and returns the output to its previous operating state.
From the remote interface:
To set the over-voltage protection for output 1 to the maximum limit:
VOLT:PROT MAX (@1)
To enable the over-current protection for output 1 and 3:
CURR:PROT:STAT ON (@1,3)
To set the over-current protection delay time for output 1 at 3 seconds:
CURR:PROT:DEL 3 (@1)
To set the over-current protection delay timer start for output 1 to CC transition:
CURR:PROT:DEL:STAR CCTR (@1)
To clear protection for output 1:
OUTP:PROT:CLE (@1)
Keysight E36300 Series User's Guide 61

Table of Contents

Other manuals for Keysight E36300 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E36300 Series and is the answer not in the manual?

Keysight E36300 Series Specifications

General IconGeneral
BrandKeysight
ModelE36300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals