3 Front-Panel Operation
52 Keysight E3633A and E3634A User’s Guide
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify
that the CV annunciator is lit. If the CC annunciator is lit, choose a higher
current limit.
– Remote interface operation:
During actual CV operation, if a load change causes the current limit to be
exceeded, the power supply will automatically crossover to the constant current
mode at the preset current limit and the output voltage will drop
proportionately.
CURRent {<current>|MIN|MAX} Set the current
VOLTage {<voltage>|MIN|MAX} Set the voltage
OUTput ON Enable the output
You can use the resolution selection keys to move the blinking digit to the right
or left when setting voltage.