EasyManua.ls Logo

Keysight E6961A - Calibrating the ENA

Keysight E6961A
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C External Instrument Calibration
62 Keysight E6961A User Guide and Method of Implementation
Figure C-2 BroadR-Reach Compliance Test Application Main Menu
2 This opens the Calibrate Disturbing Signal dialog box for the selected AWG.
3 With the appropriate AWG Address set and the correct oscilloscope channels
selected, click the Calibrate button to start the calibration process.
4 When the software finishes the calibration, click the Done button to return to
the Main Menu. Refer to the E6961A online help for a more detailed
explanation.
Calibrating the ENA
Before using the ENA, it must be calibrated using the Return Loss Calibration
board. See Figure C-3 below for the connection diagram. Calibrate the ENA using
the instructions in the ENA’s User’s Guide.

Table of Contents

Related product manuals