Keysight U1731C/U1732C/U1733C User’s Guide 53
Measuring dissipation factor/quality factor/phase angle (D/Q/θ)
The dissipation factor (D), quality factor (Q), and phase angle (θ) values can be
displayed interchangeably by pressing the key when the LCR meter is set to
the inductance, capacitance, or impedance measurement mode.
This setting is not applicable for DCR measurement.
Changing the test frequency
The test frequency is set to 1 kHz by default. Press the key to select a desired
test frequency.
Selecting parallel/series circuit mode (P/S)
The LCR meter can display parallel ( ) or series ( ) mode data for
all ranges.
Press the key for more than 1 second to toggle the parallel and series mode.
Series mode is set as the default setting. You can, however, change this power-on
behavior in the Setup menu. See “Changing the initial power-on behavior” on
page 70 for more information on how to change the default measurement mode
(parallel or series) for subsequent power cycles.
Limit
DQO
Freq.
Table 2-5 Available test frequencies
Model 100 Hz 120 Hz 1 kHz 10 kHz 100 kHz
U1731C ✔✔✔--
U1732C ✔✔✔✔-
U1733C ✔✔✔✔✔