Service ManualService Manual
Operating Lights marLEDOperating Lights marLED
®®
E9 / E9i / E15E9 / E9i / E15
V V 1.0 1.0 6969
99
Technical Data marLEDTechnical Data marLED
®®
E9 / E9i / E15E9 / E9i / E15
9.19.1
Technical Data marLED® E9 / Technical Data marLED® E9 / E9iE9i
marLED
marLED
®®
E9 E9 marLEDmarLED
®®
E9iE9i
Input voltage of power supply unit:Input voltage of power supply unit:
Output voltage of power supply unit:Output voltage of power supply unit:
Supply voltage on contact blockSupply voltage on contact block
100–240 VAC, 50–60 Hz100–240 VAC, 50–60 Hz
(28–36 VDC adjustable)(28–36 VDC adjustable)
min. 24 VDC, max. 30 VDCmin. 24 VDC, max. 30 VDC
Power Power rating rating of of power power supply supply unit unit Input: Input: 340 340 VA, VA, Output: Output: 290 290 VA VA continuous continuous operationoperation
PPoowweer r ccoonnssuummppttiioon n oon n 222200––22440 0 VVAAC C BBoooosstt:: mmaaxx. . 11002 2 VVAA
PPoowweer r ccoonnssuummppttiioon n oon n 110000––11330 0 VVAAC C BBoooosstt:: 880 0 VVAA
PPoowweer r ccoonnssuummppttiioon n oon n 2244VVAAC C BBoooosstt:: 445 5 WW
narrow narrow light light field, field, brightness brightness level level 8: 8: 35 35 WW
Power Power consumption consumption camera camera 8 8 VA VA / / 5 5 WW
Light Light sources sources High-performance High-performance LEDs, LEDs, clustered clustered in in 9 9 Light Light EnginesEngines
Service Service life life of of the the LEDs LEDs 40.000 40.000 hh
Max. Max. light light intensity intensity Ec Ec 120.000 120.000 lx lx 160.000 160.000 lxlx
Total irradiation powerTotal irradiation power
at max. light intensityat max. light intensity
ca. ca. 375 375 W/m² W/m² ca. ca. 469 469 W/m²W/m²
Photometric Photometric radiation radiation equivalent equivalent 320 320 lm lm / / W W 320 320 lm lm / / WW
Light field diameter D10,Light field diameter D10,
small–largesmall–large
16–23 16–23 cm cm 16–23 16–23 cmcm
Light field diameter D50Light field diameter D50
10 10 cm cm 9,5 9,5 cmcm
Color temperatureColor temperature
4,300 K4,300 K
Color rendering index (CRI)Color rendering index (CRI)
9595
Shadow dilution with one mask*Shadow dilution with one mask*
46 46 % % 63 63 %%
Shadow dilution with two masks*Shadow dilution with two masks*
44 44 % % 48 48 %%
Shadow dilution in standard tube *Shadow dilution in standard tube *
99 99 % % 99 99 %%
Shadow dilution in standard tube withShadow dilution in standard tube with
one mask *one mask *
45 45 % % 50 50 %%
Shadow dilution in standard tube withShadow dilution in standard tube with
two masks *two masks *
44 44 % % 45 45 %%
DimmabilityDimmability
30–100 % (backLite 5 %)30–100 % (backLite 5 %)
OperationOperation
Sterile when using sensoGrip (optional)Sterile when using sensoGrip (optional)
Non-sterile when using control panelNon-sterile when using control panel
Illumination Illumination depth depth L1+L2 L1+L2 at at 20 20 % % 330 330 + + 500 500 mm mm 330 330 + + 500 500 mmmm
at at 60 60 % % 180 180 + + 230 230 mm mm 180 180 + + 230 230 mmmm