60 Chapter 8 Troubleshooting
When you turn ON your RVG
sensor, it blinks blue then
becomes solid blue but the sensor
cannot be found in the sensor list
on the computer.
• WiFi access point is
not ON.
• Configuration
problem.
• Check that the WiFi access point is
ON.
• If it is ON and the problem persists
contact your representative
When you turn ON your RVG
sensor, it blinks blue then
becomes solid blue but when you
press the control box button to
acquire an image it does not turn
green.
The KDIS is not open Launch the KDIS
When you turn ON your RVG
sensor, it blinks blue then
becomes solid blue but when you
press the control box button to
acquire an image it does not turn
green.
The RVG sensor is not
locked on any device.
Lock the RVG sensor on the computer or
the iPhone®/iPod® touch device.
The IPS interface is frozen.
The IPS aiming ring
connection problem.
• Check that the IPS aiming ring is ON
(the shut down time is 90 sec.).
• The IPS aiming ring is too far from the
RVG sensor.
• The IPS ailing ring is to close to the
RVG sensor.
When you turn ON your RVG
sensor, it blinks blue for a long
time never becoming solid blue or
solid red.
Contact your representative.
When you turn ON your RVG
sensor, it blinks blue then rapidly it
becomes solid red.
Contact your representative.
Table 2 Quick Trouble Shooting (Continued)
Malfunction Possible Cause Action