64 Chapter 9 Maintenance
Cleaning the RVG Sensor Control Box
The RVG sensor control box must be cleaned with disposable cleaning wipes similar to those used for
the computer screens.
.
Cleaning the Positioning Accessories
.
Monthly
Wipe the outer covers of the WiFi AP with a soft and dry cloth.
Table 4 Positioning Accessories Daily Maintenance Tasks
Accessories Maintenance Tasks
Toothbrush holders
1. Remove any residue with hot water and soap.
2. Put the metal and plastic parts in separate sterilization pouch and
autoclave up to 132°C (273° F)
before the next patient.
Bite blocks
RINN Arm & ring
1. Disassemble the metal arm and the plastic ring.
2. Remove any residue with hot water and soap.
3. Put the metal and plastic parts in separate sterilization pouch and
autoclave up to 132°C (273° F)
before the next patient.
IPS aiming ring
Wipe clean
with disposable cleaning wipes.
WARNING: Never immerse the RVG sensor control box in any solution.
WARNING: Do not use chemical autoclave for the toothbrush holders
and avoid direct contact with the metallic part of the autoclave.