EasyManua.ls Logo

Konica Minolta CHROMA METER - Selecting the Close-Up Lens

Konica Minolta CHROMA METER
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Selecting the Close-up Lens
[Meas. Cond.] -
[Close-up Lens]
To measure an extremely small surface, you can use the optional close-up lens.
For the procedure to attach the close-up lens, refer to the instruction manual of the close-
up lens.
When the close-up lens is attached, it is necessary to correct the measured value by
multiplying it by the transmittance of the lens. Since the correction value varies depending
on the lens attached, you need to set the lens type on the instrument.
Setting the wrong lens type prevents proper measurement.
The following table shows the relationship between the close-up lens model and the setting
on the instrument.
Attached lens
No lens
attached
Close-up lens
No. 153
Close-up lens
No. 135
Close-up lens
No. 122
Close-up lens
No. 110
Setting None No. 153 No. 135 No. 122 No. 110
* Factory setting: None
Operating Procedure
1
5 3
2
3, 4
1. While the measurement screen is
displayed, press the
MENU
key.
The setting screen is displayed on the LCD
screen.
2. Press the
RIGHT
key or
LEFT
key to
display the [Meas. Cond.] screen.
42
Settings
OBJECTIVE
LENS

Table of Contents

Related product manuals