Setting Automatic Operation Options (Nightsmart
TM
/ SmartStart
TM
)
The cabinet allows configuration to activate functions automatically when the sash is
opened or closed. The following screens will display sequentially with [OK/Mute].
The first screen provides the option of activating the
blower; if you want the cabinet blower to start every
time you raise the sash, select Blower On, and then
[OK/Mute]. If Blower Off is selected, the blower
must be manually started from the keypad. When
[OK/Mute] is pressed, the next configuration screen
will appear.
If you want the cabinet blower to run slowly,
maintaining reduced airflows every time you close
the sash, select Reduced Speed and then
[OK/Mute]. If Blower Off is selected, the blower will
stop when the sash is closed. When [OK/Mute] is
pressed, the next configuration screen will appear.
If you want the cabinet lights to turn on every time
the sash is opened, select Light On, and then
[OK/Mute]. If Light Off is selected, the lights must
be manually illuminated from the keypad. When
[OK/Mute] is pressed, the next configuration screen
will appear if the cabinet is configured for a UV lamp.
If your Cabinet is configured for a UV light, you will
see Fig. 6-14 and 6-15. If you want the UV lamp to
turn on every time the sash is closed, select
UV Light On, and press [OK/Mute]. If UV Light Off is
selected, the UV light will not turn on when the sash
is closed. When [OK/Mute] is pressed, the final
configuration screen will appear.
OPEN SASH:
START
BLOWER
CLOSE SASH:
BLOWER
REDUCED
OPEN SASH:
LIGHT ON
CLOSE SASH:
UV LIGHT ON
Whether the UV Light is initiated from the [UV
Light] button on the keypad, or automatically
initiated upon closing of the sash, this screen
controls the time the UV lamp will remain on.
Select the length of time desired, press
[OK/Mute].
Figure 6-11
Figure 6-12
Figure 6-13
Figure 6-14
Figure 6-15
Continuous On