62
Center Work Surface Removal
To hinge the center work surface open, reference Figures 7-1 and 8-2 for part
identification and follow these steps.
1. Turn the cabinet blower on.
2. Lift either wing work surface approximately 2 inches (50 mm) by grasping the
knobs at either front corner.
3. This will lift the center work surface. Grasp the front edge of the center work
surface, and lift the front edge vertically, pivoting the work surface on its back
edge until the front edge of the work surface engages the latch on the back wall
of the work area.
If removing the center work surface from the cabinet, the work surface must be
thoroughly decontaminated before removing it from the cabinet.
To reinstall the center work surface, lift up on the latch on the back wall, and gently
lower the work surface back into position. Take care to engage the pins in the front
corners of the work surface with the holes in the grille.
Figure 8-2
LH Wing
RH Wing
Center Work
Surface
RH Wing Knob

Table of Contents