4. Position the cord or cable as it will be used in the cabinet, and then push the plug
back into the body of the portal until it seats in the portal. Reinstall the grommet.
Working in the Cabinet
Working in the Axiom Type C1 biosafety cabinet requires preparation, planning, and an
understanding of aseptic technique to work safely. The steps below represent the
minimum requirements before, during, and after working in a biosafety cabinet. Consult
your Health & Safety department for additional SOP requirements.
Planning
1. Thoroughly understand procedures and equipment required before beginning
work.
2. Arrange for minimal disruptions, such as room traffic or entry into the room while
the cabinet is in use.
Cabinet Start-up
1. Turn off UV light (if included on your cabinet).
2. Slowly raise the sash until the bottom of the sash aligns with the sash indicator
decal located on the left side of the work area (See Figure 7-1).
3. Turn on the light and cabinet blower if the automatic
features have not been
enabled.
4. Check the air grilles for obstructions.
5. Allow the cabinet to operate until the Home Screen is shown.
6. Wash hands and arms thoroughly with germicidal soap.
7. Wear appropriate personnel protective equipment (PPE).
Wipe Down
1. Raise the sash to its full open position. Mute the alarm by pressing [OK/MUTE].
2. Wipe down the interior surfaces of the cabinet with 70% ethanol, or a suitable
disinfectant, and allowed to dry.