Type A HEPA Filter Replacement
CAUTION: The cabinet must be properly decontaminated before servicing the
supply HEPA filter. Following replacement of a HEPA filter, the cabinet must
be recertified.
CAUTION: Always remove the exhaust HEPA filter first. The exhaust filter pan
or plenum can be damaged if the plenum is raised with the Exhaust HEPA
filter in place.
CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate
lifting techniques to remove and handle it. Use two people to remove the
larger filters if possible.
To Remove the Exhaust HEPA Filter on 4-, 5- and 6-foot models:
1. Unplug the cabinet.
2. Remove the front dress panel as shown in Figure 6-1.
3. Loosen all of the Phillips screws on the blower chamber cover. Remove all of the
screws except two screws on the top center of the panel. This will support the panel.
4. Release the damper seal by pushing the outside edge of the seal towards the bolt
and inside the blower plenum cover, as shown in Figures 6-14 and 6-15.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals