Product Service 1-800-522-7658
15
Quickstart
How do I run a diagnostic on the sash position switches?
1. Press the
Menu
button.
2. Press the button until the
Services
option is highlighted (it will
turn yellow).
3. Press
OK/Mute
to enter the Services menu screen.
4. Press the button until the Diagnostic
Diagnostic
option is
highlighted (it will turn white).
5. Press
OK/Mute
to enter the Diagnostic submenu.
6. Press the button until the Diagnostic
Sash sensors
option is
highlighted (it will turn white). Press
OK/Mute
.
7. Lower the sash, and note the status line it should properly
report the sash as open too high, or open, or closed when the
sash is in the corresponding position.* Press
OK/Mute
to exit
the diagnostic.
*Early production models of Type A Logic+ units may display a sash height alarm screen
when the sash is raised above its operating position. This is normal.
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics

Service
or to choose,
OK to select, MENU to escape
Sash sensors
Keypad -Relays
Canopy Alarm
Airflow Sensor

Diagnostics
OK to return
Lower the sash from full open
to closed
Note the sash status below:
Sash Status:
The sash is open
The sash is open too high
The sash is closed
Diagnostics

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals