IV Bar Kit No. 3858600, -01, -02, -03
Completely decontaminate all interior work surfaces of the Purifier before
beginning the installation of the IV bar.
Installation Procedure
1. Disconnect power to the Logic+ and raise the sash to the fully open position.
2. Remove the four screws on the sidewalls and save them.
3. Mount one of the IV brackets on either sidewall. Orient and secure the bracket
using the flat washers, lock washers and the screws as shown in Figure 1.
4. For the 5' and 6' IV Bars only, insert the center support brace on the IV Bar.
Position the center support as shown below while completing Step 5.
5. Slide the other bracket onto the IV bar, and fasten it to the opposite sidewall as
done in Step 3.
6. For the 5' and 6' IV Bars only, fasten the center support brace onto the back wall
with the two screws provided in the kit.
7. Secure each of the brackets by tightening all of the fasteners.