48
Electrical Requirements
The biosafety cabinet models have the following electrical requirements:
Model #
Requirements
All 3-& 4-foot Type A ending in -00
115 VAC, 60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -20
100 VAC, 50/60 Hz, 12 Amps
All 3-& 4-foot Type A ending in -10,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All Type B2 ending in -00
115 VAC, 60 Hz, 12 Amps
All Type B2 ending in -20
100 VAC, 50/60 Hz, 12 Amps
All Type B2 ending in -10, 30, 40,
50, 60, 70
230 VAC, 50/60 Hz, 6 Amps
All 5-& 6-foot Type A ending in -00
115 VAC, 60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -20
100 VAC, 50/60 Hz, 16 Amps
All 5-& 6-foot Type A ending in -10,
30, 40, 50, 60, 70
230 VAC, 50/60 Hz, 8 Amps
Note: A dedicated outlet with an appropriate circuit breaker should be
located as close as possible to the right rear side of the cabinet, at a
height even with, or higher than, the top of the bench or stand. Consult
your local electrical codes for properly rated circuit breakers. For safe
operation the dedicated outlet must provide the protective earthing
ground connection to the cabinet.
Note: On 100 and 115 VAC models, both electrical outlets are protected by a
ground fault interrupter circuit (GFIC). Labconco does not recommend plugging
the biosafety cabinet into a GFIC outlet.
Do not use any detachable power cord that is not adequately rated for
the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
Table 2-1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals