Product Service 1-800-522-7658
95
Vibration (optional)
The vibration level should be measured as described in ANSI/NSF Standard 49:2012.
Note: The vibration meter should have a range of 20-20,000 Hz, and have an
output in displacement.
1. Vibration level should be recorded from a single point, on the geometric center
of the work surface.
2. Take a background reading with the cabinet blower off at the test point.
3. Start the cabinet blower, and take a vibration reading.
Acceptance
Vibration should be less than 0.0002 inches displacement when the cabinet is new.
UV intensity (optional)
Note: The UV radiometer should measure light at a wavelength of 254nm.
1. UV intensity should be recorded from a single point, on the geometric center of
the work surface.
2. Take a background reading with the cabinet blower off at the test point.
3. Close the sash, turn on the UV light, and let it warm up for at least 5 minutes.
Acceptance
The UV intensity should be at least 200 uW-cm
2
.
Most common causes for insufficient UV light levels are:
d. Malfunctioning lamp(s)
e. Old lamp(s)
f. Dusty/dirty lamp surfaces

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals