Information Center
The Information Center is an LCD display located on the right side wall at eye level.
When the blower is started, the words “please wait” will be displayed, as shown in
Figure 4-2. After approximately 90 seconds, the display will switch to normal operation.
The display provides a clock, the remaining filter life, the cabinet’s current status, inflow
and downflow velocities (if equipped with the optional airflow sensor), as shown in
Figure 4-3. In the event of an alarm, the Information Center will immediately display a
context sensitive display indicating the cause of the alarm, and possible solutions, as
shown in Figure 4-4.
The display will enter sleep mode, turning itself off, one minute after the blower is
turned off or the sash is closed.