Quickstart
How do I run a diagnostic on the Canopy Alarm?
1. Press the
Menu
button.
2. Press the button until the
Services
option is highlighted (it will
turn yellow).
3. Press
OK/Mute
to enter the Services menu screen.
4. Press the button until the
Diagnostic
option is highlighted (it
will turn white).
5. Press
OK/Mute
to enter the Diagnostic submenu.
6. Press the button until the Diagnostic
Canopy Alarm
option is
highlighted (it will turn white). Press
OK/Mute
.
7. The display shows the status of the Canopy Inlet Valve. During
normal operation, it will be open, and during a canopy alarm, it
will be closed. NOTE: The canopy door function is OPPOSITE OF
THE MAGNETIC SWITCH FUNCTION! WHEN THE INLET VALVE
IS OPEN, THE MAGNETIC SWITCH IS CLOSED.
Quickstart
How do I run a diagnostic on the Airflow Sensor?
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics

Service
Diagnostics
or to choose,
OK to select, MENU to escape
Canopy Alarm
Sash sensors
Keypad - Relays
Airflow Sensor


Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals