Troubleshooting
VOICE CONFERENCE APPLICATION
Problem
Instructor microphone not picking up sound.
Possible Solutions
Unplug and re-plug the headset to your computer
Ensure correct sound device is selected. In Voice Conference
Application main menu, select <Options>, <Select Device>.
Check volume settings in windows. Ensure microphone is
not muted.
THE DEBRIEF VIEWER
Problem
Missing video capture - If the Web-camera does not record
video for debriefing.
Possible Solutions
Check that the web-camera’s USB-cable is plugged into the
Patient Monitor PC.
Check the web-camera settings via the Profile Editor.
Ensure settings match the web-camera setup.
Ensure you use the correct profile file.
Ensure that there is only one USB web-camera connected
to the PC.
The video recordings will be stored on the PC connected
to the web camera. Ensure the computer connected to the
web-cam is available when transferring to debrief.
Problem
Patient Monitor video capture is not included in the
debriefing.
Possible Solutions
Check that the name of the Patient Monitor PC being used
corresponds with the setting in the profile in use.
MANIKIN
Problem
Unpredictable behavior
Possible Solutions
Manikin malfunctions may be caused by loose cables, tubes
or connectors. Open the torso and check if any items
appear to have become disconnected or are leaking. See
section: Manikin Setup - Opening the Torso.
In case of fluid leakage, power off the manikin and contact
Laerdal Technical Service.
Problem
Identifying a single manikin if multiple manikins in network
When checking the manikin’s pulse, the <Select Manikin>
dialog of the Instructor Application will indicate which
manikin is pulse palpated.
AIRWAY CONTAMINATION
Problem
Manikin airways have become contaminated from mouth-
to-mouth rescue breathing
Solution
Clean the outside of the manikin with manikin wipes. Clean
the inside of the oral cavity with manikin wipes. Change
the lung bladders; see section: Maintenance - Replacing Lung
Bladders. The manikin airways are not designed for mouth to
mouth rescue breathing or to be disinfected.
CHEST MOVEMENT
Problem
No chest rise on manikin.
Solution
Check that manikin is switched ON.
Check that manikin is not in sleep mode due to inactivity.
Reactivate the manikin.
Ensure that the condensation valve is closed. See section:
Opening the Condensation Valve.
Check that awRR is not set to zero on Instructor
Application (Patient Case or scenario)
Check that airway complications like maximum air resistance
or laryngospasm are not set.
Check if the internal compressor is switched off. See
section: Turning the Internal Compressor Off/On.
Check that any external compressed air source is switched
off and that the air tube is disconnected from the manikin.
The internal compressor be overheated. Wait approximately
20 minutes for it to cool down.
Chest-rise is set to bilateral (for example if ET-tube is
inserted to far into the bronchia).
Chest-rise bladder is leaking or tubing to chest-rise bladder
is twisted kinked or disconnected. Replace chest-rise
bladder if it is leaking, see Maintenance section: Replacing
Chest Rise Bladders.
Check air tubing for leakage; check that all connections are
intact. Replacing tubing section if leaking.
Shallow chest movements and the internal compressor
runs continuously. The internal compressor may be worn -
contact Laerdal Technical Service.
MANIKIN LIMBS
Problems
Lack of motion in the legs
Possible Solution
Loosen and re-adjust the hip joint nuts on the inside of the
pelvis. See section: Manikin Setup - Attaching the legs.