Troubleshooting
LUNGS
Problem
Lungs not functioning properly
Possible Solution
Check that the Airway resistance is not set to maximum in
the Instructor Application.
Open the torso and chest plate. Check that the lungs are
free to expand and are not restricted by any cables.
Check that the lung bladders are properly connected, and
that the tubes are not twisted.
Check that the lung bladders are in a horizontal position
and inserted correctly. Ensure that the lung compliance
O-rings lie between the folds of the lung bladders.
Check for flaws or ruptures in the lung bladders.
Check that the two lung compliance O-rings are fitted
correctly. Replace O-rings if they appear to be damaged.
Check that there are no obstructions inside the manikin
airways which may block air flow.
If there is no change when adjusting lung compliance,
contact Laerdal Technical Service.
If there is no change in lung resistance, contact Laerdal
Technical Service.
MECHANICAL NOISE DURING AUSCULTATION
In the Instructor Application, click <auscultation focus>.
BATTERIES
Problem
Battery life is less than 150 minutes with healthy patient and
fully charged batteries.
Possible Solutions
Batteries may be old, (recommended lifetime is 200
discharge cycles). Insert new batteries.
Compressor may not function properly; consult your local
Laerdal Technical Services Centre.
PULSES
Problem
Cannot feel pedal pulses
Possible Solutions
Skin may be too tight over pulse units – re-adjust skin and
reboot.
CLINICAL FEATURES
- PNEUMOTHORAX BLADDERS
Problem
Pneumothorax bladders are not functioning properly.
Solution
Check tubing connection at base of the bladder, to ensure
that tubing has not been disconnected.
MANIKIN SHUTDOWN
Problem
The manikin is unresponsive.
Solution
Press and hold the <ON/OFF> button for 10 sec to force
shutdown the manikin.