138 cHAPTER 6: Computer Interface Operation
Model 336 Temperature Controller
MNMXRST Minimum and Maximum Function Reset Command
Input MNMXRST[term]
Remarks Resets the minimum and maximum data for all inputs.
MODE Remote Interface Mode Command
Input MODE <mode>[term]
Format n
<mode> 0 = local, 1 = remote, 2 = remote with local lockout.
Example MODE 2[term]—places the Model 336 into remote mode with local lockout.
MODE? Remote Interface Mode Query
Input MODE?[term]
Returned <mode>[term]
Format n (refer to command for description)
MOUT Manual Output Command
Input MOUT <output>,<value>[term]
Format n, +nnnnn[term]
<output> Specifies output to configure: 1–4.
<value> Specifies value for manual output.
Example MOUT 1,22.45[term]—Output 1 manual output is 22.45%.
Remarks Manual output only applies to outputs in Closed Loop PID, Zone, or Open Loop modes.
MOUT? Manual Output Query
Input MOUT? <output>[term]
Format n
<output> Specifies which output to query: 1 - 4.
Returned <value>
Format + nnnnn[term] (refer to command for description)
NET Network Settings Command
Input NET <DHCP>,<AUTO IP>,<IP>,<Sub Mask>,<Gateway>,
<Pri DNS>,<Sec DNS>,<Pref Host>,<Pref Domain>,
<Description>[term]
Format n,n,dd,dd,dd,dd,dd,s[15],s[64],s[32],
<DHCP> 0 = DHCP off, 1=DHCP on.
<AUTO IP> 0 = Dynamically configured link-local addressing (Auto IP)
off, 1 = On
<IP> IP address for static configuration.
<Sub Mask> Subnet mask for static configuration.
<Gateway> Gateway address for static configuration.
<Pri DNS> Primary DNS address for static configuration.
<Sec DNS> Secondary DNS address for static configuration.
<Pref Host> Preferred Hostname (15 character maximum)
<Pref Domain> Preferred Domain name (64 character maximum)
<Description> Instrument description (32 character maximum)