EasyManua.ls Logo

Lakeshore 336 - Page 93

Lakeshore 336
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.7.1 Alarms 81
| www.lakeshore.com
FIGURE 5-5 illustrates the interaction between alarm setpoint and dead band in
non-latching operation. With the high alarm setpoint at 100 K and the dead band
at 5 K, the high alarm triggers when sensor input temperature increases to 100 K,
and it will not deactivate until temperature drops to 95 K. In addition, the same
5 K dead band is applied to the low alarm setpoint as well.
To setup an alarm, enter the Alarm Setup menu by pressing the Alarm key. If a latching
alarm has been activated, you will be prompted with a Reset Alarm? message. Select
No to enter the Alarm Setup menu.
Menu Navigation:
AlarmQInput (A, B, C, D)QLatchingQ(Off, On)
AlarmQInput (A, B, C, D)QDeadbandQ(see note below)
Low and High Setpoint limits are determined by the Preferred Units of the associated
sensor input.
Default: LatchingQOff
DeadbandQ1.0000 K
Interface Command: ALARM
FIGURE 5-5 Dead band example
High alarm activated
Low alarm activated
Low alarm setpoint
Temperature reading
High alarm setpoint
High alarm deactivated
100 K
95 K
55 K
50 K
Low alarm deactivated
Deadband = 5 K
Alarm latching off

Table of Contents

Related product manuals