Page 79
Electronic Copy of LTI’s Impulse User’s Manual 8
th
Edition © October 1998
previously. The two readings should be within
0.05 degrees one another.
13. Reset your power conservation interval to that
of normal usage.
Note: The additive offset procedure can be
performed with a higher degree of accuracy by
setting up two targets 15 to 30 meters (50 to 100
feet) away. The center of each target must be level
with the center of the scope lens, and they must be
facing each other. This procedure requires an
accurately aligned scope. Follow the same
procedure, but instead of placing the instrument on
a desk or table, place the instrument on a stable
platform at one end of the alignment area, and
shoot to the opposite target. These two readings
should also be within 0.05 degrees of one another.