Care and maintenance
48 M320 Ref. 10734261 / Version 03 M320 Ref. 10734261 / Version 03 49
10.2 Maintenance
X
Keep accessories away from dust when not in use, e.g. protect them using a dust cover.
X
Remove dust with a pneumatic rubber bulb and a soft brush.
X
Clean lenses and eyepieces using special optics cleaning cloths and pure alcohol.
X
Thoroughly clean the optics carrier using germicidal disinfectant after each time it is handled.
X
Protect your microscope from moisture, fumes and acids and from alkaline and caustic materials. Do not store chemicals close to the
instrument.
X
Protect from improper use. Never install other device sockets or unscrew optical systems and mechanical parts unless explicitly
instructed to do so in this User Manual.
X
Protect the microscope from oil and grease. Never oil or grease the guide surfaces or mechanical parts.
X
Remove coarse contamination using a damp disposable cloth.
X
Use disinfectants based on the following active ingredients: aldehydes, alcohols, quaternary ammonium compounds.
X
Do not use preparations based on the following: halogen-splitting compounds, strong organic acids, oxygen-splitting compounds.
X
Camera: Keep optical components clean. Clean optical surfaces using a lint-free cloth. Soak the cloth using a little methanol or glass
cleaner. Do not use alcohol.
X
Do not use ethanol or spirits.
10.2.1 Tropical environment/fungus
Leica Microsystems employs certain safety precautions in its manufacturing techniques and materials. Other preventive measures include:
X
Keep optical parts clean.
X
Use and store them in a clean environment only.
X
Store under UV light when not in use.
X
Use in continuously climate-controlled rooms only.
X
Keep moisture away and cover using a plastic cover filled with silica gel.
10.2.2 Notes on reprocessing of resterilizable products
Limitations on reprocessing
X
Observe local legal regulations when processing medical products used to treat patients who have or are suspected to have
Creutzfeldt-Jakob disease (CJK) or its variant (vCJK). Usually, these resterilizable medical products can be safely disposed of by burning.
Occupational safety and health protection
X
Observe work safety and health protection of persons responsible for processing contaminated products.
X
Current regulations of hospital hygiene and prevention of infection must be observed in the preparation, cleaning and disinfection of
the products.
10.2.3 Instructions
Workplace
X
Remove surface contamination with a paper towel.