Care and maintenance
50 M320 Ref. 10734261 / Version 03 M320 Ref. 10734261 / Version 03 51
10.3 Notes on reprocessing of resterilizable products
Recommended: reprocess a product immediately after use.
Limitation of reprocessing
Frequent reprocessing has little effects on the product. The end of the product life cycle is usually determined by wear and tear and
damage through use.
Sterilization
Permissible sterilization methods
Article No. Designation Steam autoclave
134 °C, t > 10 min.
Ethylene oxide
max. 60 °C
10180591 Clip-on handle
10428328 Rotary knob, binocular tube T
10384656 Rotary knob, transparent
10443792 Lever extension
10446058 Protective glass, multifocal lens
1)
10446469 Protective objective glass,
Leica M680/FL400
1)