Care and maintenance
50 M320 Ref. 10734261 / Version 03 M320 Ref. 10734261 / Version 03 51
10446467 Protective objective glass,
Leica M840/M841
1)
10445341 Handle for Leica M655, sterilizable
10445340 Cap for Leica M655/M695, sterilizable
10446842 Handle for Leica M400, sterilizable
10448440 Cover, sterilizable for Leica M320 handle
10448431 Protective objective glass, Leica M320
1)
10448296 Objective protective glass, Leica M720, spare part (package of
10)
1)
10448280 Protective objective glass, Leica M720, complete, sterilizable
1)
10448581 Cover, sterilizable for RUV800
1)
Products with optical components can be steam-autoclaved using the conditions listed above. However, this may cause a layer of dots
and streaks to form on the glass surface, which may reduce the optical performance.
Cleaning
Needed: water, detergent, spirits, microfiber cloth
1 Flush the surface with running water (<40 °C), using a little detergent if necessary.
2 Also use spirits to clean optical components.
3 Dry optical components using a microfiber cloth, dry the rest of the product using a paper towel.
Disinfection
X
After disinfection, thoroughly clean optical surfaces using running water/fresh drinking water and then rinse using fresh, demineralized
water.
X
Dry the products completely before the subsequent sterilization.
Leica Microsystems (Schweiz) AG validates:
The instructions above are suitable for preparing a product to be reused. The processor is responsible for the desired results. Before
deviating from the instructions provided, first verify the deviations for effectiveness and possible consequences.
Maintenance
The Leica M320 surgical microscope is maintenance-free. To ensure operational safety and reliability, Leica Microsystems (Schweiz) AG
recommends taking the precaution of contacting the responsible service organization. There, periodic inspections can be agreed or a
maintenance contract can be concluded.