Specifications
58 M320 Ref. 10734261 / Version 03 M320 Ref. 10734261 / Version 03 59
13.9 Electromagnetic compatibility (EMC)
13.9.1 Environment for which the instrument is suitable
Normal environmental conditions as well as hospital
13.10 Compliance IEC 60601-1-2
Emissions
CISPR 11, Class B, Group 1
Harmonic distortion per IEC 61000-3-2 Class A
Voltage fluctuation and flicker per IEC 61000-3-3 Class A, Figures 3-7
Immunity
Electrostatic discharge CD ±8 kV, AD ±15 kV
Radiated RF EM fields 80 - 2700 MHz: 10 V/m
Proximity wireless fields 380 - 5785 MHz: 9 V/m; 28 V/m
Electrical fast transients and bursts ±2 kV power supply lines
Surges ±1 kV differential mode; ±2 kV common mode
Conducted disturbances, induced by RF fields 10 V rms
Rated power-frequency magnetic field 30 A/m
Voltage dips and interruption according to IEC 60601-1-2:2014
Rated power-frequency magnetic field Not applicable