6 | Operation LINXON
42 / 90 LX218-Operating-Manual-en1-04-(1908)
Options Value range Comments
turbo molecular pump
(TMP) cannot be vented
during run-up.
Standby > Configuration > Global settings > Access control > Maintenance activated
Access to CAL function
Define the permission for carrying out a calibration of the device.
Options Value range Comments
Enable calibration YES The calibration can be
started via the "Standby"
window.
No The calibration cannot be
started via the "Standby"
window.
Standby > Configuration > Global settings > Access control > Access to the CAL
function
Change device PIN
The device PIN controls the use of the device.
If this function has been activated, then a personal identification number (PIN) must be
entered to use the device. Only those who know the PIN and enter it correctly can
start the device. The device can not be used without entering the correct PIN. The
device PIN is requested immediately after switching on the device. If you enter an
incorrect PIN, the following message appears: "Incorrect PIN".
Authorization for the device is active if the current device PIN is not 0000.
If you activate authorization for the device, you have to save the device PIN.
If you have forgotten the device PIN, contact Linxon.
Options Value range (min./max.) Comments
New PIN 0000 – 9999 New device PIN
New PIN (confirmation) 0000 – 9999 New device PIN. Repeat to
confirm.
Standby > Configuration > Global settings > Access control > Change device PIN
Changing the menu PIN
The menu PIN controls access to the software menu of the device. If this function has
been activated, a personal identification number (PIN) must be entered to access the
menu for the device. Only those who know the PIN and enter it correctly can access
the device menu.