LINXON Operation | 6
LX218-Operating-Manual-en1-04-(1908) 47 / 90
Options Value range (min./max.) Comments
If the leak detector is put
back in Standby/STOP
mode, a new background
is displayed after 25
seconds at the latest.
Off With START, the internal
mass spectrometer
background (MS-BG) is not
subtracted.
See description under
"On".
Standby > Configuration > Test settings > Filter & ZERO
Information on active background suppression can be found in the status line of the
measurement screen:
Display Setpoint
ZERO after pressing the ZERO button briefly in
the "activated" or "with Start" setting
ZERO Start appears after the pre-set time has
passed in the "with Start" setting
6.4.3 Setting vacuum ranges
CAUTION
Damage to turbo pump due to high load
Operation at 15 ... 25 mbar places a heavy load on the turbo pump. Avoid continuous
operation in this pressure range.
NOTICE
Changing the default settings can result in significant device performance
reduction. If you have questions, contact the manufacturer.
Standby > Configuration > Test settings > Vacuum ranges
These settings can only be set for the operation mode "Vacuum":
Options Value range (min./max.) Comments
ULTRA Activated Activated