LINXON Operation | 6
LX218-Operating-Manual-en1-04-(1908) 59 / 90
The calibration history shows the date, time, tracer gas, and calibration factor for the
last 12 calibrations. Additional details are stored for the calibrations. To view details
about a calibration, go to the path shown above.
Then select the following:
1. In order to select a certain calibration, use the buttons next to the symbols ↑ or ↓
to move the selection to the respective calibration.
2. In order to display the complete message, select the magnifying glass symbol.
Paging function
Standby > Configuration > Information > Paging function
1. Switch the paging function on or off. When the paging function is switched on, the
remote control unit will send out an audible signal.
6.7 Calibrating with vacuum method
CAUTION
For optimal calibration, the device must be warmed up for at least 20 minutes.
Please observe the recommended test interval for the calibration leak used! See
quality test result: Calibration leak.
In the vacuum mode, the device can be calibrated with an internal or external
calibration leak. The internal calibration is only possible for mass 4.
Internal calibration leak
There are two ways to calibrate with an internal calibration leak:
• Automatic internal: For calibration with the internal calibration leak without volume
at the inlet flange. For this purpose, the inlet flange must be blind-flanged.
• Manual internal: For calibration with the internal calibration leak in the presence of
a volume at the inlet flange. Confirmation of a stable measurement signal by
means of the "Signal stable" button is necessary.
External calibration leak
For a calibration leak, the following prompt appears: "External calibration: Connect
and open external calibration leak".