LINXON Operation | 6
LX218-Operating-Manual-en1-04-(1908) 51 / 90
Options Value range (Min. Max.) Comments
than the warning limit and
less than the setpoint, a
continuous sound is
output.
6.4.8 Defining the calibration settings
Standby > Configuration > Calibration settings
In this menu, you can define the settings for the calibration. The calibration is not
introduced here.
Options Value range (min./max.) Comments
Units e.g., mbar*l/s The units for the calibration
leak value. The units for
the internal calibration leak
are mbar*l/s and can not
be changed.
Calibration leak value
(internal/external)
Calibration leak value in
the selected units.
Depending on the selected
calibration mode, this is
either an external or
internal calibration leak.
Operation mode
"Calibration"
Int. auto Internal calibration mode,
automatic.
Int. man. Internal calibration mode,
manual – the signal
stability must be calibrated
manually.
External
6.4.9 Setting the calibration request
Run-up > Configuration > Calibration settings > Calibration request
Specify whether the device displays a calibration request after standard events occur.
Option Comments
YES The calibration request is carried out. –
30 minutes after switching the device on -
if the temperature in the device has
changed by more than 5 °C since the last
calibration.