2. Check if the parameters displayed in the "Standby" window are correct. The
device is ready when the "Standby" window appears on the display after the
device starts up. Make sure that the device is set to the "Sniff" operation mode.
3. In order to obtain the most precise measurement results possible, allow the device
to warm up for approx. 30 minutes.
4. To start the measuring process, press "START" on the control unit. The
measurement window is displayed.
5. In order to take the background concentration of the tracer gas into consideration
and suppress it during the measurements, keep the sniffer tip away from possible
gas sources and press the ZERO button on the device or on the button on the
handle briefly.
6. Hold the sniffer tip close to the possible leak source and sniff the test object. The
tip must touch the test object. If you would like to test a weld or something similar,
you must move the tip at a speed of less than 2.5 cm/s (1 in/s) along the path.
When checking a spot, hold the sniffer line for at least 1 second.
If there is a leak, this is indicated in the display as well as by an acoustic signal,
regardless of your settings.
6.5.3 Measurement view
Once the measuring pressure is reached, the measurement view appears with the
most recently used display type:
Analog/digital with bar graph and large format numbers or
Graphic as a function of measuring time
By using the lower right button, you can switch between the analog/digital display
and the graphical display. This button alternately shows the analog display symbol
and the graphical display symbol.
6.6 Calling up information about the device
You can call up various information about the state of the device. The information is
shown on the display in the form of lists or graphics.
Run-up > Configuration > Information
Settings
System data
Vacuum diagram
Error list
Calibration history
You can also locate the installed remote control using an acoustic signal:
Standby > Configuration > Information > Paging function