6 | Operation LINXON
44 / 90 LX218-Operating-Manual-en1-04-(1908)
6.3.1.4 Volume and beeping
WARNING
Damage to hearing due to loud audio
The alarm level of the device can exceed 85 dB(A).
► Adjust the volume accordingly.
► Only expose yourself a short time to the alarm.
► Use hearing protection.
Standby > Configuration > Global settings > Beep volume
Beeping sound: Enable or disable the sounds. The sounds signal, for example, a
change of state.
Minimum volume: You can set the minimum volume. You can set the volume in the
measurement view. The setting applies to the speaker in the device.
Setting range: 0 to 15
6.4 Settings for the measurements
6.4.1 Selecting the operation mode and mass
• Select the operation mode
• Select the gas (mass)
• Set the leak rate factor
If a menu PIN is required to access the software menu, this function is not available
until the PIN is entered.
Option Value range (min./max.) Comments
Operation mode Vacuum
Sniffing
Operation mode "Vacuum"
Operation mode "Sniffing"
Connect the sniffer line
before pressing START
Mass H
2
(2 amu)
3
He (3 amu)
4
He (4 amu)
Detectable gas H
2
Detectable gas
3
H
Detectable gas
4
H
Leak rate factor Factor
1E-6 … 1E+6
Leak rate converted using
user-defined factor