LINXON Warning and error messages | 7
LX218-Operating-Manual-en1-04-(1908) 75 / 90
No. of the
warning
message
Displayed message Description and possible remedy for the
cause
• A software update has been performed
and one or more new parameters
determined. The message can be
confirmed in this case. The parameter(s)
is (are) automatically initialized.
W152 No external pressure
sensor detected!
No external pressure sensor has been
detected.
• Ext. External measuring tube selected in
the menu, but not connected or cable is
defective
• Ext. pressure sensor defective.
Characteristic impedance not detected.
W160 Leak rate too high!
Switched to the "Standby"
mode to prevent
contamination!
The monitor function "Contamination
protection" is activated and a leak rate
above the set limit value has been detected.
• Gross leak.
• Switch-off limit value too low.
• Alarm delay set too short.
W161 Maximum evacuation time
exceeded!
Within the set evacuation time, the
"Measuring" state has not been reached.
• Evacuation time is adapted incorrectly to
the sample volume.
• DUT has a gross leak.
• Switching pressures set incorrectly.
Table8:
Warnings