Integration
“Integration” is a sub-menu allowing access to the “Enable,” “Value,”
“Multiplier,” and “Units” associated with the integration alert for View 1.
If the alert is enabled, the instrument will generate an alert message and tone
when the current integrated dose is equal to value represented by the
“Value,” “Multiplier,” and “Units” associated with the alert.
Note: The error associated with displaying the integrated dose on the
measurement view is one numerical unit. As a result, an alert value of 200 uR
may appear to generate an alert when the integrated dose is between 199.9
and 200.1 uR; however, the actual detected integrated dose was 200.0 uR.
Enable
“Enable” represents the ability to enable the integrated dose alert for the
current view. “Enable” is a special field and valid values include “Off”
and “On.”
Value
“Value” represents the numerical value of the integrated dose alert given
the supplied multiplier and units. “Value” is a numeric field with valid
values from 0 to 999.
Multiplier
“Multiplier” represents the multiplier associated with the integrated dose
alert. “Multiplier” is a special field with valid values of micro, milli, unity,
kilo, and mega.
Units
“Units” represents the units associated with the integrated dose alert.
“Units” is a non-editable field, as it must be relative to the units
associated with the current view measurement rate.
Alarms
“Alarms” is a sub-menu allowing access to the “Radiation” and “Integration”
alarms associated with “View 1: R/hr”.