MODEL 9DP-1 Operators Manual
Ludlum Measurements, Inc. 40 December 2015
Integration
Integration is a sub-menu allowing access to the Enable, “Value,
“Multiplier, and “Units” associated with the integration alert for View 1.
If the alert is enabled, the instrument will generate an alert message and tone
when the current integrated dose is equal to value represented by the
“Value, “Multiplier, and “Units associated with the alert.
Note: The error associated with displaying the integrated dose on the
measurement view is one numerical unit. As a result, an alert value of 200 uR
may appear to generate an alert when the integrated dose is between 199.9
and 200.1 uR; however, the actual detected integrated dose was 200.0 uR.
Enable
Enable represents the ability to enable the integrated dose alert for the
current view. Enable is a special field and valid values include “Off
and “On.
Value
“Value” represents the numerical value of the integrated dose alert given
the supplied multiplier and units. Value” is a numeric field with valid
values from 0 to 999.
Multiplier
“Multiplier” represents the multiplier associated with the integrated dose
alert. “Multiplier is a special field with valid values of micro, milli, unity,
kilo, and mega.
Units
“Units represents the units associated with the integrated dose alert.
“Units is a non-editable field, as it must be relative to the units
associated with the current view measurement rate.
Alarms
Alarms is a sub-menu allowing access to the Radiation and Integration
alarms associated with View 1: R/hr.